Abraxis: 522011:  Cylindrospermopsin in water by Immunoassay, Microtiter Plate

  • Summary
  • Analytes
  • Revision
  • Data and Sites
Official Method Name
Abraxis Cylindrospermopsin ELISA Plate Kit PN 522011
Current Revision
2006
Media
WATER
Instrumentation
Immunoassay
Method Subcategory
Organic
Method Source
  Abraxis
Citation
  Abraxis User Guides and Flowcharts
Brief Method Summary
Cylindrospermopsin is detected using a colorimetric immunoassay (ELISA) procedure. A sample (0.05 mL), enzyme conjugate (Cylindrospermopsin-HRP), and an antibody solution against Cylindrospermopsin are added to plate wells pre-coated with sheep anti-rabbit antibodies. Both the Cylindrospermopsin (if present) in the sample and Cylindrospermopsin-HRP conjugate compete to bind to the antibody solution in proportion to their concentrations. The anti-Cylindrospermopsin antibodies are then bound to the immobilized sheep anti-rabbit antibodies on the plate. After incubation, the unbound molecules are washed and decanted. A substrate is then added which is catalyzed by the enzyme and converted from a colorless to a blue solution. The reaction is terminated with the addition of a dilute acid. The concentration of Cylindrospermopsin in the sample is determined by measuring its absorbance at a specific wavelength (450nm) using a photometer and comparing its absorbance to the absorbance of the calibrators.
Scope and Application
This method determines Cylindrospermopsin in water (groundwater, surface water, well water).
Applicable Concentration Range
0.04 - 2.0
Interferences
No interferences known
Quality Control Requirements
(A) Calibration with 6 standards and 1 blank, all analyzed in duplicate. (B) Precision: 3 matrix samples with different levels in the range for quantitative analysis analyzed daily for 5 days with 5 replicates in each of 5 assays. (C) Accuracy: 5 matrix samples spiked with the target analyte at 4 different levels in the range for quantitative analysis. (D) Validation: Analysis of 4 positive and 4 negative samples by an independent method, for confirmation.
Sample Handling
Samples are collected in glass containers with Teflon-lined caps; for compliance monitoring, samples should be held no longer than 14 days. If samples are held frozen, they can be held for longer periods of time. Immunoassay reagents are stored refrigerated until use. Samples containing gross particulate matter should be filtered using a 1 micron filter or centrifuged before use. Drinking water samples are typically dechlorinated with 0.008% sodium thiosulfate.
Maximum Holding Time
14 days at 4 C, longer if held frozen
Relative Cost
Less than $50
Sample Preparation Methods