Standard Methods: 2510B: Conductivity - Laboratory Method
|
Official Method Name
|
Conductivity - Laboratory Method |
|---|---|
|
Current Revision
| Standard Methods Online |
|
Media
|
WATER |
|
Instrumentation
|
Resistor Network |
|
Method Subcategory
|
Inorganic |
|
Method Source
|
|
|
Citation
|
|
|
Brief Method Summary
|
In the laboratory, conductance (or resistance) of a standard KCl solution is measured and from the corresponding conductivity, a cell constant is calculated. Once the cell constant has been determined, or set, the conductivity of an unknown solution will be displayed by the meter. |
|
Scope and Application
|
This method covers the determination of conductivity in surface waters, drinking water, and wastewater. |
|
Applicable Concentration Range
|
None given. |
|
Interferences
|
None Provided. |
|
Quality Control Requirements
|
Calibration of instrument |
|
Sample Handling
|
Collect at least 500 mL sample in a polyethylene or glass bottle. Refrigerate. |
|
Maximum Holding Time
|
28 days |
|
Relative Cost
|
Less than $50 |
|
Sample Preparation Methods
|
None required |