Modern Water: A00111/A00128:  PCP (pentachlorophenol) in soils/sediment by immunoassay

  • Summary
  • Analytes
  • Revision
  • Data and Sites
Official Method Name
Pentachlorphenol (PCP) RaPID Assay A00111, with PCP Soil Extraction Kit A00128 (SDI 1998) http://www.sdix.com/
Current Revision
1998
Media
SOILS/SEDIMENT
Instrumentation
Immunoassay
Method Subcategory
Organic
Method Source
  Modern Water
Citation
  Modern Water Inc. User Guides
Brief Method Summary
Inhibition immunoassay, with a colorimetric indicator for competitive binding of analyte and an enzyme-labeled conjugate to surface-immobilized antibodies. Colorimetric response is inversely related to the (log) concentration of analyte in the sample or standard. Calibrated for PCP. Reagents are purchased in kits with manufacturers' instructions for sample preparation and analysis. Requirements: a supply of clean water and methanol, to extract soil and sediment samples, to dilute sample extracts, and to prepare standards; filtration equipment to remove particulates from water samples and soil extracts; an oven and an analytical balance, to dry soil and sediments after extraction, and accurately determine sample weights; an SDI magnetic bead separator, required for the SDI RaPID assay kit; a spectrometer, which accepts sample tubes (SDI), for analyses at visible wavelengths (450 nm); equipment to automate sample preparation for large numbers of samples (prefiltration, measurement of uL volumes, rinsing, agitation), so that contact times with reagents are controlled as required; computerized data acquisition and processing. Expendable supplies - approximately $8 per determination
Scope and Application
Ambient and compliance monitoring. Can be used for Method 4010A. Soil screening for PCP by immunoassay (EPA 1995).
Applicable Concentration Range
0.10 - 10 ug/g
Interferences
Sensitivity < PCP: 2,3,5,6- and 2,3,4,6- tetrachlorophenol; 2,3,5- and 2,3,6-trichlorophenol; tetrachlorohydroquinone, Sensitivity << PCP: 2,4,6- , 2,4,5- , and 2,3,4-trichlorophenol; 2,4- , 2,6- , 2,3- , 2,4- , and 3,5-dichlorophenol; hexachlorobenzene; hexachlorocyclohexane
Quality Control Requirements
Recommended for a kit allowing 100 quantitative determinations:

a. calibration with 3 standards and 1 blank, all analyzed in duplicate

b. precision: all 46 samples analyzed in duplicate (repeats)

c. accuracy: 3 matrix samples, spiked with the target analyte at different levels in the range for quantitative analysis.

d. validation: analysis of 4 positive and 4 negative samples by an independent method, for confirmation
Sample Handling
Samples are typically collected in glass containers with Teflon-lined caps. Samples are held in a refrigerator at 4 oC. Immunoassay reagents and kits are stored in the dark at 4 to 8 oC until use.
Soil and sediment samples need to be dried, weighed, and extracted (methanol; clean water). Soil extracts generally need to be centrifuged or prefiltered to 0.2 um. Quantitative analysis requires dilution of highly concentrated samples, until they are bracketed by standards in the linear response range. Aliquots of sample extracts and reagents are accurately measured (uL) and mixed for required times, following manufacturer's instructions for use of the materials purchased.
Maximum Holding Time
5 days at 4oC until extraction and analysis
Relative Cost
Less than $50
Sample Preparation Methods