EPA-NERL: 239.1: Lead by Flame AA
|
Official Method Name
|
Lead (Atomic Absorption, Direct Aspiration) |
|---|---|
|
Current Revision
| Issued 1971; Editorial Revision 1974 and 1978 |
|
Media
|
WATER |
|
Instrumentation
|
Flame Atomic Absorption |
|
Method Subcategory
|
Inorganic |
|
Method Source
|
|
|
Citation
|
|
|
Brief Method Summary
|
A sample prepared according to EPA Method 200.0 is analyzed using a flame atomic absorption spectrometer. Specifications for instrument operation and preparation of standards are provided in the method. |
|
Scope and Application
|
This method determines lead by flame atomic absorption spectrometry. Sample pretreatment according to EPA Method 200.0 determines the applicability. |
|
Applicable Concentration Range
|
1-20 mg/L (optimal range); 0.5 mg/L (sensitivity). |
|
Interferences
|
For interferences see EPA Method 200.0. Sensitive to turbulence and absorption bands in the flame. Flame alignment and signal noise must be centered and quieted as much as possible. |
|
Quality Control Requirements
|
QC requirements for drinking water procedures are provided in method 200.0 |
|
Sample Handling
|
Sample handling and preservation procedures are provided in EPA Method 200.0. |
|
Maximum Holding Time
|
6 months (MCAWW, Table 1). |
|
Relative Cost
|
Less than $50 |
|
Sample Preparation Methods
|
200.0 |