EPA-NERL: 252.2: Osmium by Graphite Furnace AA
|
Official Method Name
|
Osmium (Atomic Absorption, Furnace Technique) |
|---|---|
|
Current Revision
| Issued 1978 |
|
Media
|
WATER |
|
Instrumentation
|
Graphite Furnace-Atomic Absorption Spectrometer |
|
Method Subcategory
|
Inorganic |
|
Method Source
|
|
|
Citation
|
|
|
Brief Method Summary
|
A sample prepared according to EPA Method 200.0 is analyzed using a graphite furnace atomic absorption spectrometer. Specifications for instrument operation and preparation of standards are provided in the method. |
|
Scope and Application
|
This method determines osmium by graphite furnace atomic absorption spectrometry. Sample pretreatment according to EPA Method 200.0 determines the applicability. |
|
Applicable Concentration Range
|
50-500 ug/L (optimal range). |
|
Interferences
|
Background correction is recommended. Since this metal volatilizes near 150oC, the allowable ashing temperature must be verified in the sample matrix being analyzed. For additional interferences see EPA Method 200.0. |
|
Quality Control Requirements
|
QC requirements for drinking water procedures are provided in method 200.0 |
|
Sample Handling
|
Sample handling and preservation procedures are provided in EPA Method 200.0. Since osmium tetroxide is very volatile and highly toxic, great care should be exercised when working with this compound. |
|
Maximum Holding Time
|
6 months (MCAWW, Table 1). |
|
Relative Cost
|
$51 to $200 |
|
Sample Preparation Methods
|
200.0 |